Search results for "Blood Glucose Measurement"
showing 2 items of 2 documents
Impact of physical exercise on sensor performance of the FreeStyle Libre intermittently viewed continuous glucose monitoring system in people with Ty…
2019
AIMS To evaluate the sensor performance of the FreeStyle Libre intermittently viewed continuous glucose monitoring system using reference blood glucose levels during moderate-intensity exercise while on either full or reduced basal insulin dose in people with Type 1 diabetes. METHODS Ten participants with Type 1 diabetes [four women, mean ± sd age 31.4 ± 9.0 years, BMI 25.5±3.8 kg/m2 , HbA1c 55±7 mmol/mol (7.2±0.6%)] exercised on a cycle ergometer for 55 min at a moderate intensity for 5 consecutive days at the clinical research facility, while receiving either their usual or a 75% basal insulin dose. After a 4-week washout period, participants performed the second exercise period having sw…
Evaluation of CGMS® During Rapid Blood Glucose Changes in Patients with Type 1 Diabetes
2006
CGMS (Medtronic Minimed, Duesseldorf, Germany) allows continuous glucose monitoring. Recent studies with invasive monitoring techniques revealed discrepancies in blood glucose measurements obtained from different anatomical sites compared with those from the fingertip. The aim of this study was to investigate the CGMS and a device for alternative site testing (AST) during dynamic blood glucose changes and to compare these results with fingertip measurements.Twelve patients with type 1 diabetes (seven women, five men; age, 33.3 +/- 8.7 years) received a 75-g oral glucose load. Insulin was applied intravenously (rapid glucose decline) or subcutaneously (moderate glucose decline) in a dosage b…